Microdialysis and microperfusion electrodes in neurologic disease monitoring

Abstract

Contemporary biomarker collection techniques in blood and cerebrospinal fluid have to date offered only modest clinical insights into neurologic diseases such as epilepsy and glioma. Conversely, the collection of human electroencephalography (EEG) data has long been the standard of care in these patients, enabling individualized insights for therapy and revealing fundamental principles of human neurophysiology. Increasing interest exists in simultaneously measuring neurochemical biomarkers and electrophysiological data to enhance our understanding of human disease mechanisms. This review compares microdialysis, microperfusion, and implanted EEG probe architectures and performance parameters. Invasive consequences of probe implantation are also investigated along with the functional impact of biofouling. Finally, previously developed microdialysis electrodes and microperfusion electrodes are reviewed in preclinical and clinical settings. Critically, current and precedent microdialysis and microperfusion probes lack the ability to collect neurochemical data that is spatially and temporally coincident with EEG data derived from depth electrodes. This ultimately limits diagnostic and therapeutic progress in epilepsy and glioma research. However, this gap also provides a unique opportunity to create a dual-sensing technology that will provide unprecedented insights into the pathogenic mechanisms of human neurologic disease.

Introduction

The United States Center for Disease Control (CDC) estimated in 2015 that 1.2% of the American population has active epilepsy, including approximately 3 million adults and half-a-million children [1]. Conservatively estimating that 20% of such patients have medically refractory epilepsy, this leaves roughly 700,000 patients in need of neurosurgical procedures or other novel strategies to achieve disease control [2]. Notably, CNS neoplasms comprise the underlying seizure focus in a subset of patients with intractable epilepsy. While some CNS neoplasms are benign, indolent, or surgically curable, most primary CNS malignancies remain incurable. Indeed, the most common adult primary brain tumor is glioblastoma (GBM), which is typically fatal in just over a year from diagnosis [3], with under 5% of patients surviving 5 years [4]. The development of new treatments for glioma has remained disappointing, with no new drug shown to improve survival since temozolomide was introduced in 2005 [5]. Therefore, at present, the therapeutic toolbox for both epilepsy and glioma are inadequate.

Paramount in the immediate surgical management of both epilepsy and glioma is optimally balancing the preservation of neurocognitive function while simultaneously achieving maximal safe resection of epileptogenic or tumor-infiltrated brain tissue. However, some gliomas and epileptic foci cannot be effectively removed without unacceptable loss of neurocognitive function. In such cases, neurosurgical avenues may still be pursued for diagnostic or therapeutic procedures, spanning biopsy, electrocorticography from subdural or penetrating electrodes (ECoG), stereotactically placed brain-penetrating electrodes (sEEG), and, less frequently, focal CNS delivery of therapies via convection-enhanced perfusion or chronic implantation of stimulation-based devices to mitigate propagation of seizures.

Human neurological diseases are notoriously challenging to accurately model in vitro or in preclinical models. However, collection, identification, and quantitation of extracellular biomarkers from the diseased human brain may offer the potential to enhance our understanding of pathogenic mechanisms and thereby accelerate the development of much-needed therapies [6]. Neuronal activity has recently been shown to facilitate glioma progression [7], but at the same time certain electrical stimulation paradigms attenuate tumor growth by increasing chemotherapeutic efficacy [8]. Moreover, seizure activity may both result from and lead to neurochemical and metabolic aberrations in the brain [6]. As such, the coordinated collection of electrical and biochemical information in these diseases may expose novel cause-and-effect relationships. The capture of correlated or interdependent neurochemical and electrical fluctuations may yield hypotheses relevant to the treatment of both epilepsy and glioma. Moreover, since electrodes can both record and stimulate, and since microdialysis or microperfusion can both sample and focally introduce agents into the extracellular biochemical environment, the combination of these modalities robustly expands the breadth of therapeutic hypotheses that can be directly tested in a patient-specific manner explicitly within diseased human CNS tissue.

Since biochemistry and electrical activity are the currencies of neurophysiology, applications of purpose-built technologies could extend beyond epilepsy to include patients undergoing neurosurgical procedures for stroke, traumatic brain injury, Parkinson disease, and other neurological and neurodegenerative conditions. Since effective technologies for concurrent biochemical and electrical interaction within the CNS are a prerequisite to progress in this arena, we address the key engineering-related factors of relevance to developing optimal dual-purpose devices, first addressing microdialysis and microperfusion separately before reviewing precedent efforts to combine these with invasive EEG recording electrodes.

Comparisons between microdialysis and microperfusion

Microdialysis and microperfusion are similar biomarker collection techniques that comprise a porous partition between the surrounding tissue microenvironment and perfusate flowing within the sampling device, allowing extracellular parenchymal compounds to diffuse down a concentration gradient across the partition [9] (Fig. 1). Many system designs include at least one precision pump responsible for propelling biomarker-free perfusate or microdoses of pharmaceutical candidates [10] from a reservoir to the diffusion probe where dialysis occurs [11]. The biomarker-rich perfusate, consequently termed dialysate, then continues to a microvial for stabilization and storage, prior to identification and quantitation of compounds such as pharmacological molecules, neurotransmitters, antibacterials, cytotoxic agents, cytokines, and metabolites [9]. Small molecules, peptides, and proteins are characterized in the dialysate using biochemical analysis, enzyme-linked immunosorbent assay, or liquid chromatography and mass spectrometry [12, 13].

Fig. 1
figure 1

Microdialysis and microperfusion systems. The major features of microdialysis and microperfusion devices include cylindrical probes, reservoirs for metabolites and biomarkers, pumps to drive perfusion fluid flow and control fluid fluctuation into the tissue. Exclusive to microdialysis is a membrane that separates and protects the peri-probe tissue from the flowing perfusate. Microperfusion incorporates a secondary pull pump to draw perfusate from the catheter and perfusion ports, allowing larger molecules and compounds to diffuse into the collection fluid 

A primary strength of these technologies is the ability to continuously collect perfusate using minimally invasive implantation [9] of probes that permit the analysis of biomarkers over the course of events such as electrical stimulation or during systemic or localized administration of pharmacological candidates [14,15,16,17]. Such continuous collection is also regularly used to monitor the effects of the implanted probe on the biochemistry of the peri-probe tissue [18,19,20,21]. An additional strength of microdialysis and microperfusion is the potential to streamline the determination of effective drug dosing, especially for compounds such as antibiotics, through localized dosing at micro levels [12]. Microdialysis and microperfusion provide a platform for baselining an analyte within a targeted area while inducing no net biological fluid or tissue loss, a significant advantage over the use of surgical resection for tissue sampling [22]. However, both sampling approaches are plagued by skewing of biomarker collection induced by tissue damage during implantation [20, 23,24,25,26], and both approaches require complicated calibration techniques to accurately determine recovery rate, defined as the percentage of collected biomarker relative to the true tissue concentration [27].

Microdialysis and microperfusion in contrast

Microdialysis techniques progressed substantially in the 1990s and have been validated repeatedly thereafter, although with decreasing frequency in the last 10–15 years, providing foundational data and reliability as a developed science [9, 11, 14,15,16, 28]. The most significant distinctive quality of microdialysis in contrast to microperfusion is the incorporation of a semipermeable membrane to control the selection of sought after biomarkers based upon molecular weight cutoff (MWCO), measured in Daltons [21]. At any designated MWCO, roughly 80–90% of the molecules at that particular molecular mass are unable to transverse through the membrane while molecules of lower mass are able to pass through the membrane in a graded manner [9] (Fig. 2). The exact percentage of exclusion versus permeability for any specific molecular weight, however, varies between membrane manufacturers and does not behave like a step function. Also, the MWCO specified by the manufacturer may not provide an accurate prediction of the experimental recovery rate, even if the targeted compound has a molecular weight precisely at the MWCO. Selecting a membrane with a MWCO that is considerably larger than the molecular weight of the compound of interest is recommended to ensure measurable quantities of the compound are sampled experimentally [29]. However, while high molecular weight molecules such as cytokines, growth factors, and large neuropeptides have been successfully captured using high MWCO membranes [30], the recovery rates remain unpredictable. For example, the estimated recovery of cytokines across a 100 kDa MWCO membrane is only 1–5% at normal flow rates, despite these factors having molecular weights that are generally under 25 kDa [31, 32]. A unique solution to this challenge is to change the effective gradient “drive” for the cytokine across the membrane by adding polymer microbeads coated with antibodies to the perfusate. This strategy has been shown to increase the relative recovery of some cytokines in vitro by upwards of 20-fold [31] and may be a viable method for increasing recovery of many proteins in vivo.

Fig. 2
figure 2

Microdialysis membrane molecular weight cut-off. Recovery rate across a defined MWCO membrane can be approximated by an exponential function that decreases as the molecular weight of the sampled compound increases [137, 138]. Mechanistically, this is related to the presence of progressively fewer large-diameter pores in the membrane relative to small-diameter pores, resulting in reduced probability of large molecule permeation. The largest pore size sets the absolute cut-off for permeability, but the effective cut-off is considerably smaller. The manufacturer specified MWCO for a membrane is a single point along a spectrum where compounds of a known molecular weight experience a specific transmission rate across the membrane, typically 10-20% (or conversely, 80-90% retention or permeability resistance). Therefore, to achieve high recovery (>80%) of a desired molecule with a specific molecular weight (MW-A), a membrane with a much larger MWCO must be employed (MWCO-B)

Other benefits of microdialysis include the relative absence of tissue damage or hydrodynamic disruption after implantation is achieved. In contrast to microperfusion, microdialysis membranes can isolate tissue structures from possibly detrimental hydrodynamic forces associated with perfusate flow and therefore can induce less overall tissue damage [10]. Finally, microdialysis setups typically require only one push pump to achieve adequate perfusion, due to the physical retention of perfusate within the probe preventing flow into the peri-probe tissue [33] (Fig. 1).

Conversely, microperfusion avoids the need for the dialysis membrane and maximizes biomarker collection via the use of open collection ports in the indwelling catheter. The indiscriminate pore structure of the microperfusion probe circumvents the reduced recovery rates associated with microdialysis membranes (Fig. 2) and overcomes failure to collect sought after biomarkers due to unpredicted molecular interactions between the analyte of interest and the membrane. These ports, typically less than 1 mm in diameter, allow compounds of any size to fluctuate into the perfusate [34] (Fig. 1). Microperfusion probes are more commonly developed in experimental labs and cleanrooms [35] than microdialysis probes due to the simplicity of creating fluid channels compared to complex microdialysis membranes. Several groups have developed unique probes in-house to evaluate the design efficacy of stereolithography and micromachining [36], droplet generators [35], and soft lithography [37] for probe construction. Others laser etch holes into existing catheters to simplify the fabrication process [38]. However, microperfusion probes require a secondary pull pump to ensure a zero net fluctuation of fluid from the probe into the interstitial space [33, 34, 39]. Open flow microperfusion (OFM) has been validated in several studies in human skeletal muscle [40], adipose tissue [40], and cerebral tissue (cOFM) [21], supporting the validity of the approach in vivo.

As described, both microdialysis and microperfusion have distinct advantages (Table 1). Microdialysis is a highly documented sampling technology that excels for the collection of molecules below a predetermined molecular weight using a one pump system [33]. Unfortunately, the difficulty in finding appropriate commercial membranes along with membrane biofouling often hinder experimental success [26]. Conversely, microperfusion employs a simpler probe construction to sample molecules of any size [33] but has less of a published presence, prolonging experimental procedures and slowing technical implementation since the technology is still developing. Additionally, microperfusion requires a pull pump as well as a push pump to adequately control fluid flow [33].

Table 1 Advantages and disadvantages of microdialysis and microperfusion

Microdialysis and microperfusion probe design parameters

The most critical design parameter in the microdialysis system is the semipermeable membrane (Fig. 2), responsible for establishing diffusion of the targeted biomarker across a gradient and for maintaining a net-zero fluid volume fluctuation [27]. Similarly, the size and quantity of open ports that comprise the microperfusion diffusion area are the critical part of the microperfusion probe [36]. Critical microdialysis membrane characteristics include construction material, surface charges that inhibit compound adhesion and dampen the interaction of the membrane with the neuroinflammatory response, MWCO, and more intuitive parameters such as geometry, temperature, stability, and reusability [41].

A plethora of membrane materials have been developed and successfully implemented in preclinical and clinical studies, including poly(carbonate-ether), cuprophan, and poly(acrylonitrile) [42]. Other membrane material options include poly(carbonate), poly(arylethersulphone), poly(ethersulphone), poly(urethane), or cellulose [10, 43]. The membrane material properties can critically impact sampling efficiency for biomarkers such as peptides, due to charge interactions [44], or proteins, due to absorption on the membrane [45].

Apart from material, the total area where diffusion occurs as well as inlet/outlet positioning are also critical determinants of performance for both microdialysis and microperfusion. Larger diffusion areas improve recovery rate by naturally providing more surface area for concentration fluctuation, although the data is variable and dependent upon target compound [12]. Microfabricated probes can suffer from low recovery rates due to reduced surface area and minor variations in the fabrication process [36]. Inlet and outlet geometry can be configured using a concentric tube design or a side-by-side configuration according to fabrication limitations, and material resources, while the material, such as fluorinated ethylene propylene, polyimide, stainless steel, or fused silica can be tailored to considerations such as dielectric or stiffness requirements or susceptibility to MRI fields [20, 21].

Microdialysis and microperfusion performance parameters

Flow rates in each technology are commonly in the 0.5–1 µL/min range [31, 46] but have been experimentally investigated down to 0.25 µL/h (~0.004 nL/min) [47]. The basic premise guiding flow rate decisions is that recovery rate or extraction coefficient is inversely related to volumetric flow rate [27, 31, 46].

Additionally, lower volumetric flow rates decrease the probability of tissue damage induced by impinging fluid forces as the perfusate passes the collection site [10]. Computational modeling indicates the potential for higher shear stresses in microdialysis as compared to microperfusion [48]. With regard to microperfusion, evidence suggests that ultralow flow rates induce minimal tissue damage [49]. There is evidence that ultralow microperfusion flow rates (less than 10 nL/min) induce localized neuroinflammatory responses, though reducing the flow rate even further, to less than 5 nL/min, was associated with a reduced response [47].

Recovery rate calibration

The concentration of the targeted compound in the collected solution does not necessarily represent the concentration of that same compound in the tissue, because diffusion varies depending on the surrounding tissue properties and flow properties internal to the probe [9]. Despite the 80–90% nominal membrane resistance estimate at the defined membrane MWCO, calibration is necessary to accurately estimate actual tissue concentration using both microdialysis and microperfusion [40, 50]. It is important to recognize that the measured concentration of the collected biomarker is often much less than the “true” concentration in the brain because of the combined diffusion resistance of the dialysate, the membrane, and the tissue [27]. The relationship between the concentration in the dialysate, \({C}_{D}\), and the tissue, \({C}_{T}\), is mathematically characterized by the recovery rate, RR, and is dominated by the flow rate, \(Q\), and resistances to diffusion, \(\sum R\), as shown in the following equations [51]:

$$RR=\frac{{C}_{D}}{{C}_{T}}.$$
$$RR=1-exp\left[\frac{-1}{Q(\sum R)}\right].$$

Total resistance to diffusion, R, can be expressed as a series function:

$$\sum R={R}_{D}+{R}_{M}+{R}_{E},$$

where RD is the resistance of the dialysate, RM is the resistance of the membrane, and RE is the resistance of the surrounding tissue or environment [51]. In vivo resistance is a dynamic and rapidly changing function of tissue metabolism, cellular and vasculature exchange, and true diffusion [51]. In vitro calibration values cannot be extrapolated to in vivo samples due to the behavior of the live tissue relative to a synthetic media [51]. Therefore, it is necessary to calibrate the recovery rate of microperfusion and microdialysis experiments to calculate the true concentration of the targeted compound in the surrounding media. The no-net-flux calibration approach, also termed the equilibrium method, along with the reverse dialysis method [9] and extrapolation to zero method [12, 52] are together the most common calibration techniques and are further explored below. However, these calibration methods are subject to adhesion of the target analyte(s) to the flexible cannula between the membrane and the collection reservoir and to other plastic surfaces involved in the closed system. This non-linear loss must be taken into consideration for accurate calibration and it is important that the field works toward a more thorough understanding of how analytes (ranging from small molecules to large protein complexes and vesicles) adhere to the device. The development of active anti-fouling and anti-adhesion probes is a critical need.

In vivo no net flux calibration

The no-net-flux calibration method exploits repeated dialysate measurements while adding and incrementing the concentration of the targeted substance of choice into the dialysate inlet [40]. Measurement of the concentration is taken after the supplied dialysate is subjected to the unknown gradient between itself and the tissue [22]. After plotting the difference between input and outlet concentrations against the inlet concentration, the point at which the input concentration equals the outlet concentration (x-intercept) signifies the absolute concentration of the molecule in the peri-probe tissue, while the slope of the regression indicates the recovery rate [22, 40] (Fig. 3). A transient no-net-flux technique was also developed to decrease calibration duration for time-sensitive protocols and clinical trials [53].

Fig. 3
figure 3

No-net-flux calibration method. Plotting the experimentally determined differences between output (Cout = dialysate) and input (Cin = perfusate) concentrations for a molecule of interest against the known input concentrations yields an estimate of actual tissue concentration at the point where the linear regression crosses the abscissa. In addition, the slope of the regression provides the recovery rate. While time consuming, this method provides an accurate estimate of tissue concentration with minimal a priori assumptions

Retrodialysis calibration

The reverse dialysis method, termed retrodialysis, internal standard, or delivery method, has been well validated in vivo [22, 27, 46]. This procedure assumes that diffusion of any given molecule is equivalent in both directions across a membrane [22]. The perfusate is “spiked” with a known concentration of the analyte and the outlet (dialysate) concentration is measured and then used to back calculate the recovery rate using the following equation:

$$Recovery \,\, Rate \left(\%\right)= 100-\left(\frac{{C}_{Dial}}{{C}_{Perf}}*100\right).$$

A critical point for this calibration method is that diffusion is only equivalent in both directions if the initial concentration of the analyte in the peri-probe tissue prior to calibration is zero [22]. The presence of analyte in the peri-probe tissue prior to calibration will yield an underestimated recovery rate and an overestimated absolute concentration. This aspect makes the retrodialysis calibration method useful for measurement of exogenous drug levels but limited for measurement of endogenous biomarkers.

Extrapolation to zero calibration

The extrapolation to zero method is executed by plotting the flow rate against the measured in vivo concentration over several different incremental flow rates [22]. The extrapolated intercept represents the concentration at a zero flow rate and therefore the absolute concentration of the substance of interest in the peri-probe tissue, with:

$${C}_{Dial}= {C}_{o}(1-{e}^{\frac{-rA}{F}}).$$

where CDial is the measured concentration of the analyte, Co is the concentration of the substance around the probe, r is the mass transfer coefficient, A is the surface area of the membrane, and F is the experimentally defined flow rate [54]. Due to the nature of the exponential characterizing function, multiple very slow flow rates are recommended to minimize extrapolation error [22].

These calibration methods, although more popular in microdialysis, are also applicable to microperfusion, showing less than 10% variability between extrapolation to zero method and the no-net-flux method [34].

Invasive consequences of implantation and biofouling

Published literature indicates several critical parameters to reduce tissue damage during probe insertion and post-implantation that will improve strategies for successful biomarker sampling. However, despite efforts to eliminate tissue damage, the nature of probe implantation is such that damage can only be mitigated, not prevented [18, 25]. For example, reduced analyte recovery [26] and acute changes in electrophysiological properties [16] as a consequence of probe implantation can be detected within minutes. In parallel, the neurochemical properties of potentially damaged tissue immediately adjacent to the probe differ from that of similar tissue as close as 1 mm away [25]. Implantation-induced contamination of the peri-probe tissue with blood-borne inflammatory factors is also a critical limitation for the accurate assessment of inflammatory cytokines and chemokines using either microdialysis or microperfusion [23, 55]. Mitigation of this issue may be realized by allowing the blood–brain barrier to heal prior to sampling, provided the experiment can be prolonged by up to two weeks [19, 21, 23]. Previous work has shown that the blood–brain barrier is a dynamic interface for the transportation of selective compounds [56], therefore emphasizing the need for the barrier to heal prior to sampling. If healing is not possible due to time constraints, then additional sampling may be necessary in order to isolate the influence of the barrier. For example, simultaneous testing in the blood, at the barrier, and in the brain, per se.

Factors that skew the collection of biomarkers due to tissue damage are the size [48, 57], geometry [58], and rigidity [59] of the implanted probe. The volume of damaged tissue increases with the cross-sectional size of the probe [39]. The cross-sectional geometry of the probe also plays a role in the variety and population of collected biomarkers [58,59,60]. Abnormally elongated cell nuclei are present directly adjacent to a cylindrical shaft but not to the thin edge of the same custom probe [58], and other probe geometries induce less neuronal loss along thin edges as compared to traditional cylindrical probes [58, 60]. Finally, a flexible polyimide electrode probe has been shown to limit the inflammatory response [59], suggesting that construction of flexible microdialysis and microperfusion probes may reduce tissue damage and improve accuracy and reliability.

Biosensor failures can also stem from biocompatibility failures such as biofouling [26]. The membrane biofouling layer that forms on biosensors tends to be avascular and fibrous, reducing analyte recovery gradually over time, in contrast to the acute effect mediated by tissue damage during probe implantation, which improves with time [26]. Implanted microdialysis probes have been shown to induce progressively increasing reactivity in both astrocytes and microglia in the probe vicinity over the course of a month, indicating biofouling rather than an invasive trauma response [20]. Biomimicry and microperfusion flow show strong potential for reducing biofouling due to the performance of biomimicry in the first 2 weeks post implantation [26] and 97% retained sensitivity of microperfusion flow technology in the first four hours after implantation [26].

Not only do the invasive consequences of probe implantation and biofouling affect the efficacy of biomarker collection, these factors may also impact underlying pathogenic mechanisms directly. Tissue sites surrounding microdialysis probes have shown elevated production of cytokines such as interleukin-1β (IL-1β) within one hour of implantation [61], IL6 within two days [62], and elevated astrocyte reactivity within one week [62]. These factors may suppress tumor growth or exacerbate peritumoral injury [63]. They may also promote neuronal function and synaptic maintenance [64,65,66] or elicit network dysfunction and seizures [67].

Electrode design and performance

There are several design and performance parameters to consider in neural depth electrode development. Depth electrodes are categorically broken into microelectrode designs capable of capturing EEG data from single neurons [68, 69] and macro electrodes that collect signal over several square millimeters [70]. Microelectrodes are regularly fabricated through lithography techniques including sputtering or electron beam (e-beam), chemical vapor deposition, and etching [71,72,73] to capitalize on electrode density over small surface areas, although those processes can be extrapolated for macroelectrodes, as well [74]. For both electrode types, conductor materials tend to be platinum iridium [74], especially in patient studies [75,76,77], or titanium [78, 79], tungsten [80, 81], or biocompatible magnesium [72] surrounded by insulative polymers such as polyimide [73, 82, 83] and PDMS [72, 75].

Microelectrodes have benefits in precision data collection and the reduction of tissue response to implantation, just as miniature microdialysis and microperfusion probes reduce inflammatory responses that correlate with probe size [84]. Tissue disruption is in proportion to the cross sectional area of the probe [81]. After implantation, microelectrodes can be implanted to target specific cell layers [76] or provide localized data from sub-millimeter scale neuronal assemblies [77]. Multiple microelectrodes can monitor spatiotemporal patterns unlike a single macroelectrode [76] and can increase spatial density due to the nature of their size. Microelectrode implantation techniques may increase surgical risk to the patient while decreasing the accuracy of implantation [76, 85]. Lesion volume does not appear different between microelectrodes and macroelectrodes [85].

Conversely, macroelectrodes can provide similar data in terms of cellular spiking by delineating areas of cellularity [76, 86]. Macroelectrodes provide larger activation areas that linearly decrease radially away from the electrode in terms of stimulation [76]. Microelectrodes stimulate more cells relative to their occupied volume with activation profiles that decrease more rapidly away from the probe [76]. Macroelectrodes are preferred for long term implantation due to improved durability [86].

In terms of electrode performance, corrosion of electrodes and concomitant surface area reduction increases current densities and chemical reactivity and therefore skews data over time [87]. Regardless of electrode type, work has identified factors to improve biocompatibility and reduce corrosion and therefore collect more consistent data. Boron-doped diamond electrodes have been developed to improve electrochemical properties and improve corrosion resistance over several months [79, 88,89,90,91]. Platinum electrodes are recommended over stainless steel electrodes to maintain biocompatibility and minimize risk of signal drift [87]. Plating tungsten microelectrodes with platinum black coatings reduces impedance and improves post-stimulation cell activation [92].

Modifications of the insulation surrounding the electrodes can also improve electrode functionality and biocompatibility. Specifically, lowering polyimide curing temperatures leads to more stable chronic data collection [83]. Surface adhesion of biomolecules and polyurethane hydrogels promote biocompatibility by reducing protein absorption [93] and astrocyte attachment [94]. Lastly, reducing micromotion of implanted probes via probe geometry and flexibility decreases the magnitude of chronic gliosis [73].

Previously fabricated microdialysis and microperfusion electrode systems

Several microdialysis and microperfusion electrodes have been designed and tested to tackle the in vivo sensing challenge of simultaneously measuring changes in biomarkers during passive electrophysiological recording or active electrical stimulation of brain tissue [16, 95,96,97,98,99,100,101,102,103,104,105]. Some of the more elementary designs involve two separate probes for neurochemical sampling and EEG recording [95, 96, 102, 104, 106,107,108,109,110,111,112,113,114]. Other designs successfully incorporate both technologies into a single probe [98,99,100,101, 103] or even fix separate probes together through various means [115,116,117,118,119,120,121]. The majority of these studies were conducted in vitro or in non-human animal models, with minimal evidence obtained from clinical studies to date [122, 123].

Preclinical examples of dual-sensing systems employing microdialysis and microperfusion electrodes

The default setup for concurrently measuring neurochemicals along with EEG is commercially purchased microdialysis probes located separately from implanted stimulating/recording electrodes [96, 102, 104, 106,107,108,109,110,111,112,113,114]. Separate microdialysis and electrochemistry probes have been deliberately placed in rat striatal locations at least 2 mm apart to determine local gradients of dopamine and dopamine metabolites following amphetamine administration [102]. Most preclinical efforts measure dopamine or glutamate in rat brains where the probe separation distances are unspecified [106,107,108, 110,111,112]. Multi-electrode arrays have been used to measure local field potentials while simultaneously collecting extracellular analytes via microdialysis in opposing mouse brain hemispheres [113]. Another experiment measured EEG at the cortical surface [114] while simultaneously collecting small molecule biomarkers and metabolites by microdialysis in other locations. Lastly, one group fabricated and effectively used a microperfusion probe to measure glutamate fluctuations induced by a stimulating electrode in the opposing hemisphere of an anesthetized mouse [35].

To minimize the spatial gaps between the sEEG probe and the neurochemical sampling probe, several studies have employed a commercially available microdialysis cannula epoxied to an empty guide cannula to permit simultaneous collection of small molecules in proximity to a concentric bipolar electrode in the rodent caudate and nucleus accumbens targeting GABA and other amino acids [115, 119,120,121]. Others have attached recording electrodes directly to the microdialysis cannula to collect zinc ion [116, 117] or phenytoin for monitoring seizure activity [118]. A critical early study utilizing this methodology employed a flow-through hollow fiber with 50 kDa MWCO epoxied to a bipolar electrode to measure changes in taurine, glutamate, and glycine during seizures induced by intrahippocampal infusion of quinolinate [124]. To further minimize positional differences between the microdialysis probe and the electrodes, some setups incorporate conventional microdialysis with a concentric electrochemical electrode to measure neurotransmitters in two different modalities [125, 126].

Clinical examples of dual-sensing systems employing microdialysis and microperfusion electrodes

Clinically, cerebral microdialysis has been performed with a hybrid invasive EEG electrode in neurosurgical patients with medically intractable epilepsy to delineate resectable seizure foci. In a seminal study from During and Spencer [122], a microdialysis probe was attached to a polyurethane/silastic brain-penetrating electrode (referred to as a depth electrode or stereo-EEG (sEEG) electrode) which was inserted into the hippocampus of patients with refractory epilepsy. In conjunction with subdural strip electrodes, this study captured changes in glutamate and GABA after electrophysiologically defined seizures. In another critical study [123], flexible polyurethane invasive, tissue-penetrating electrodes (Ad-Tech) with 6 or 7 platinum macroelectrodes spaced along the shaft and four to nine 40 μm diameter platinum/iridium microwires inserted within the lumen extending 4-5 mm beyond the probe tip were used for electrophysiological recording. A 200 μm diameter cuprophan microdialysis membrane (Akza Nobel Faser AG) was inserted through the same lumen and extended 10 mm beyond the tip of the electrode, revealing changes in multiple amino acids during seizures and cognitive tasks. While various other clinical strategies have been employed for dual electrophysiology and neurochemical sensing in humans, the intralumenal microdialysis-penetrating electrode configuration remains the state-of-the art [127].

Relevant patents and commercial systems

A US patent from 1995 [98] claims a dialysis electrode device in which a semipermeable membrane forms the outer wall of a hollow probe while a working electrode, reference electrode, and counter electrode are fed through the top of the probe head, shielding the functional electrode from the tissue environment [98]. However, this device is not truly a dual sensing platform, as the perfusate is not under flow during measurement and the system uses an enzymatic electrode that can only assess one analyte at a time. Another US patent, from 1994 [128], claims a more generalized approach in which a cylindrical microdialysis probe ensheaths an internal “primary probe” that may collect electrophysiological data. Even more generally, a patent from 2003 claims a multi-lumen catheter system that can simultaneously support multi-modal interrogation of tissues, including the brain [129]. A search for patents incorporating “microdialysis”, “brain”, and “electrode” failed to identify any specific inventions that collect dialysate under flow either via microdialysis or microperfusion coupled with simultaneous collection of electrophysiological data in a form factor that is attentive to the size and geometry limitations discussed above.

A commercial version of the intra-lumen microdialysis-depth electrode system, employed by Spencer et al. [127], was manufactured by Ad-Tech with the explicit purpose of temporally synchronizing microdialysis, invasive EEG, and single neuron recordings. A standard Spencer depth electrode with 8 platinum macro-electrodes and 8 platinum micro-electrodes was modified such that 4 rows of perforations spaced 90 degrees circumferentially were placed between two sets of micro- and macro-electrodes at a location between approximately 4 mm and 20 mm from the electrode tip. The electrode was further modified at a point distal to the recording area (extracranially located after insertion of the electrode) such that 20 kDa or 100 kDa MWCO microdialysis catheters could be inserted (M Dialysis 70 and 71 brain catheters). This configuration allowed macroscopic perfusion through the perforations so that extracellular fluid was freely in contact with the microdialysis membrane within a region straddled by recording electrodes. Unfortunately, for unknown reasons, this device is no longer available from Ad-Tech.

Discussion and open challenges

While microdialysis and microperfusion are closely related in construction, configuration, and purpose, the benefits of microperfusion generally outweigh the advantages of microdialysis. Microperfusion provides a simpler design allowing end-users to avoid limited commercial options for MWCO, thereby yielding versatility in probe geometry and perfusion port design through in-house fabrication that often requires minimal overhead equipment. Microperfusion casts a wider biomarker net and supports improved sensitivity to large compounds that have difficulty diffusing across the high-resistance dialysis membrane [31], but cannot restrict molecules of a particular molecular weight that might reduce signal-to-noise in the analyte analyses. Notably, microperfusion can collect biomarkers to reveal preclinical and clinical phenomena that would otherwise not pass through the semipermeable membrane of microdialysis. Such novel biomarkers include extracellular vesicles and other membrane-bounded structures as well as multi-molecular complexes and aggregates.

Calibration is required and should be conducted in vivo using the no-net-flux method if time permits, or via the extrapolation to zero method, but likely not the reverse dialysis method, assuming that the compound of interest exists in the peri-probe tissue. Despite the utility of these calibration techniques under well-controlled conditions, application of these methods within the dynamic clinical setting is challenging. This is due to insufficient time to vary parameters, limited time to achieve steady-state conditions, and concerns regarding introduction of exogenous materials into the host tissue. In practice, such conditions cannot be achieved in patients and current calibration methods are inadequate to improve estimates under non-steady state conditions. Aside from the relative fluctuations in collected biomarkers, application of microdialysis and microperfusion under real-world clinical conditions – especially within the context of the CNS – will require the development of novel techniques to provide accurate estimates of absolute tissue concentrations. Both microdialysis and microperfusion provide effective ways to measure biomarker fluctuations over the course of time of epileptic events but fall short in terms of measuring absolute concentrations in the clinical setting. Furthermore, the relationship between neurochemical fluctuations and invasive electrophysiology (local field potentials, multi-unit and single neuron activity) measurements is experimentally characterized by a lag due to the time that it takes compounds to diffuse through semipermeable brain tissue [51] as well as the time it takes electrophysiology signals to be transmitted and processed. Although the lag from collecting neurochemical samples can be several minutes [130], mathematical models exist for estimating molecular diffusion rates in brain tissue [131] and for correlating with electrophysiology signal processing [132, 133]. Additionally, data collected from microdialysis and microperfusion studies can be compared to the limited data of previous clinical studies as a means of validation [15]. Future microperfusion and microdialysis studies will benefit from additional published data sets correlating neurochemical biomarkers to brain electrophysiology in both healthy and diseased patients using traditional methods.

Inflammatory and reactive responses to the probe remain problematic despite the reduction of biofouling in the microperfusion realm [26]. While the inflammatory response skews acute data collection [16, 26] and biofouling reduces recovery over time [26], it is also evident that these responses affect the tumor microenvironment and induce seizure-like discharges. The presence of cytokines, especially factors such as IL-1β, near the probe implantation site have been shown to support immune reaction against the growth of tumor in other studies [63], suggesting a benefit of probe implantation in the vicinity of a tumor. Conversely, factors such as IL-1β and TNFα have the potential to directly drive aberrant discharges [134] and seizures [67] and may therefore exacerbate disease in patients with refractory epilepsy. Probe designs that emphasize minimally invasive techniques may mitigate some of the influence of inflammatory and tissue injury factors on the recovery of biomarkers. Regardless of how minimally invasive an implantation technique is, however, inflammatory tissue responses remain unavoidable without the development of active anti-inflammatory or anti-reactivity strategies. Successful mitigation steps also include implementation of miniaturized [57] flexible [59] or anchored [73] probes with flow rates beneath the threshold for perfusate pressure-induced tissue damage [135]. Ultimately, the widespread use of clinically-implanted catheters will be hampered until probe designs are developed that do not induce significant injury and inflammatory responses in the brain [59].

In terms of invasive electrophysiology electrodes, penetrating EEG electrodes are largely broken down into microelectrodes and macroelectrodes. Microelectrodes provide enhanced ability to monitor localized neuronal populations and single neuron activity compared to large-area macroelectrodes that sample the collective activity of large neuronal populations [77, 136]. Macroelectrodes may also provide more durability for long-term monitoring [86]. For both types, platinum-iridium electrodes show strong evidence of tissue compatibility [87, 122, 123] and cleanroom fabrication processes can promote further biocompatibility.

Several studies have been conducted to collect neurochemical biomarkers concurrently with invasive EEG data. Incorporating multiple probes, especially commercially purchased probes, can expedite experimental progress at the expense of collecting data that overlooks localized neurochemical fluctuations. Most of these published experiments have been conducted in rodents while utilizing commercially-available microdialysis probes in separate locations from the brain-penetrating electrodes [96, 102, 104, 106,107,108,109,110,111,112,113,114] despite evidence that extracellular biomarker diffusion is severely limited by tissue resistance [51, 110]. Therefore, it is crucial that the neurochemical sampling probe is spatially coincident with the target neurons to properly correlate fluctuations in analytes with electrophysiology data. Most studies performed in rodent models have not captured such physically coincident analyte and electrophysiological data. The setups that successfully collected neurochemical data and electrophysiology data from the same probe employed small-molecule microdialysis and targeted neurotransmitters such as GABA and glutamate and other amino acids [115, 119,120,121] with improved spatial resolution [115]. These preclinical setups did collect relatively coincident data correlating neurotransmitters to seizure activity but were limited by the lack of sensitivity to large compounds due to the nature of microdialysis.

Clinically, the studies mentioned above utilized microdialysis in conjunction with invasive electrophysiology electrodes to correlate seizure activity to GABA, glutamate, and other small molecules [122, 123]. The separation distances between the location of analyte sampling and EEG sampling were not well specified [122] and could be upwards of 10 millimeters depending upon implantation deflection [123]. As with the preclinical experiments, the focus was mostly limited to seizures and associated neurochemicals, with no evidence regarding larger molecules. Ultimately, the simultaneous capture of large molecules and electrical fluctuations may yield hypotheses relevant to the treatment of both epilepsy and gliomas. The combination of these modalities will robustly expand the breadth of therapeutic hypotheses that can be directly tested in a patient-specific manner explicitly within diseased human CNS tissue.

Despite the plethora of previously developed and published microdialysis electrodes and microperfusion electrodes, no known microperfusion electrode system has been developed for preclinical or clinical use. Thus, there is a significant unmet need for the development of novel microperfusion electrode dual-sensing probe designs that can concurrently collect large neurochemical biomarkers such as cytokines, metabolites, cell-free DNA, extracellular vesicles, and microRNA at the precise location of brain electrophysiology measurements (single neuron, multi-neuron, local field potentials, and EEG) monitoring. Coupled with high resolution electrophysiological monitoring spatially coincident with microperfusion ports, such a system will allow unprecedented insight into the dynamic environmental changes associated with neural function and dysfunction.

Conclusion

Microdialysis and microperfusion are sampling techniques that may provide unique insights into neurologic disease microenvironments, especially when paired with invasive electrophysiology recordings. While both microdialysis and microperfusion have separate advantages, far more work has been done to validate preclinical and clinical use of microdialysis, leaving microperfusion as a relatively unexplored technology that may ultimately have more clinical potential and advantages. Calibration of both approaches is well documented and best conducted through the No-Net-Flux method, if conditions permit. Concerns regarding the tissue inflammatory response and biofouling may be addressed with new geometries and the development of active, rather than passive, anti-inflammatory properties built into the probe materials. Multiple previous endeavors have incorporated invasive EEG and neurochemical sampling into localized areas by fastening separate probes or designing one-off microfluidic electrodes to target dopamine and other neural biomarkers, mostly in small animal experiments and in vitro settings. The lack of adequate techniques for collecting neurochemical compounds such as cytokines and other large molecules, the absence of clinical in vivo validation testing of microperfusion electrodes, and the existence of large spatial separations between electrodes and perfusate collection locations results in a significant unmet gap between the available technology and the devices needed to support novel advances in epilepsy and GBM research.

Availability of data and materials

Not applicable.

References

  1. Epilepsy Data and Statistics 2019. Available from: https://www.cdc.gov/epilepsy/data/index.html.

  2. Kwan P, Brodie M. Early identification of refractory epilepsy. N Eng J Med. 2000;342:314–9.

    Article  CAS  Google Scholar 

  3. Sottoriva A, Spiteri I, Piccirillo SGM, Touloumis A, Collins VP, Marioni JC, et al. Intratumor heterogeneity in human glioblastoma reflects cancer evolutionary dynamics. Proc Nat Acad Sci. 2013;110(10):4009–14.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  4. Ostrom QT, Bauchet L, Davis FG, Deltour I, Fisher JL, Langer CE, et al. The epidemiology of glioma in adults: a “state of the science” review. Neuro-Oncology. 2014;16(7):896–913.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. van Genugten JAB, Leffers P, Baumert BG, Tjon-a-Fat H, Twijnstra A. Effectiveness of temozolomide for primary glioblastoma multiforme in routine clinical practice. J Neuro-Oncol. 2010;96(2):249–57.

    Article  CAS  Google Scholar 

  6. Bianchi L, De Micheli E, Bricolo A, Ballini C, Fattori M, Venturi C, et al. Extracellular levels of amino acids and choline in human high grade gliomas: an intraoperative microdialysis study. Neurochem Res. 2004;29(1):325–34.

    Article  CAS  PubMed  Google Scholar 

  7. Venkatesh Humsa S, Johung Tessa B, Caretti V, Noll A, Tang Y, Nagaraja S, et al. Neuronal activity promotes glioma growth through neuroligin-3 secretion. Cell. 2015;161(4):803–16.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Janigro D, Perju C, Fazio V, Hallene K, Dini G, Agarwal MK, et al. Alternating current electrical stimulation enhanced chemotherapy: a novel strategy to bypass multidrug resistance in tumor cells. BMC Cancer. 2006;6(1):72.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Sottoriva A, Spiteri I, Piccirillo SGM, Touloumis A, Collins VP, Marioni JC, et al. Intratumor heterogeneity in human glioblastoma reflects cancer evolutionary dynamics. Proc Nat Acad Sci. 2013;110(10):4009-14.

    Article  CAS  PubMed  Google Scholar 

  10. Bourne JA. Intracerebral microdialysis: 30 years as a tool for the neuroscientist. Clin Exp Pharmacol Physiol. 2003;30(1–2):16–24.

    Article  CAS  PubMed  Google Scholar 

  11. Joukhadar C, Derendorf H, Muller M. Microdialysis a novel tool for clinical studies of anti-infective agents. Eur J Clin Pharmacol. 2001;57(3):211–9.

    Article  CAS  PubMed  Google Scholar 

  12. Hutchinson PJ, O’Connell MT, Al-Rawi PG, Maskell LB, Kett-White R, Gupta AK, et al. Clinical cerebral microdialysis: a methodological study. J Neurosurg. 2000;93(1):37–43.

    Article  CAS  PubMed  Google Scholar 

  13. Emmett MR, Andrén PE, Caprioli RM. Specific molecular mass detection of endogenously released neuropeptides using in vivo microdialysis/mass spectrometry. J Neurosci Methods. 1995;62(1):141–7.

    Article  CAS  PubMed  Google Scholar 

  14. Reinstrup P, Ståhl N, Mellergård P, Uski T, Ungerstedt U, Nordström C-H. Intracerebral microdialysis in clinical practice: baseline values for chemical markers during wakefulness, anesthesia, and neurosurgery. Neurosurgery. 2000;47(3):701–10.

    CAS  PubMed  Google Scholar 

  15. Mellergård P, Åneman O, Sjögren F, Pettersson P, Hillman J. Changes in extracellular concentrations of some cytokines, chemokines, and neurotrophic factors after insertion of intracerebral microdialysis catheters in neurosurgical patients. Neurosurgery. 2008;62(1):151–8.

    Article  PubMed  Google Scholar 

  16. Jay TM, Zilkha E, Obrenovitch TP. Long-term potentiation in the dentate gyrus is not linked to increased extracellular glutamate concentration. J Neurophysiol. 1999;81(4):1741–8.

    Article  CAS  PubMed  Google Scholar 

  17. Lynes J, Jackson S, Sanchez V, Dominah G, Wang X, Kuek A, et al. Cytokine microdialysis for real-time immune monitoring in glioblastoma patients undergoing checkpoint blockade. Neurosurgery. 2018;84(4):945–53.

    Article  PubMed Central  Google Scholar 

  18. Wisniewski N, Klitzman B, Miller B, Reichert WM. Decreased analyte transport through implanted membranes: differentiation of biofouling from tissue effects. J Biomed Mater Res Ban. 2001;57(4):513–21.

    Article  CAS  Google Scholar 

  19. Birngruber T, Ghosh A, Perez-Yarza V, Kroath T, Ratzer M, Pieber TR, et al. Cerebral open flow microperfusion: a newin vivotechnique for continuous measurement of substance transport across the intact blood–brain barrier. Clin Exp Pharmacol Physiol. 2013;40(12):864–71.

    Article  CAS  PubMed  Google Scholar 

  20. Birngruber T, Ghosh A, Hochmeister S, Asslaber M, Kroath T, Pieber TR, et al. Long-term implanted cOFM probe causes minimal tissue reaction in the brain. PLoS ONE. 2014;9(3):e90221.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Birngruber T, Sinner F. Cerebral open flow microperfusion (cOFM) an innovative interface to brain tissue. Drug Discov Today Technol. 2016;20:19–25.

    Article  PubMed  Google Scholar 

  22. Plock N, Kloft C. Microdialysis—theoretical background and recent implementation in applied life-sciences. Eur J Pharmaceut Sci. 2005;25(1):1–24.

    Article  CAS  Google Scholar 

  23. Ghosh A, Birngruber T, Sattler W, Kroath T, Ratzer M, Sinner F, et al. Assessment of blood–brain barrier function and the neuroinflammatory response in the rat brain by using cerebral open flow microperfusion (cOFM). PLoS ONE. 2014;9(5):e98143.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Groothuis DR, Ward S, Schlageter KE, Itskovich AC, Schwerin SC, Allen CV, et al. Changes in blood–brain barrier permeability associated with insertion of brain cannulas and microdialysis probes. Brain Res. 1998;803(1–2):218–30.

    Article  CAS  PubMed  Google Scholar 

  25. Khan AS, Michael AC. Invasive consequences of using micro-electrodes and microdialysis probes in the brain. TrAC Trends in Analytical Chemistry. 2003;22(8):503–8.

    Article  CAS  Google Scholar 

  26. Wisniewski N, Reichert WM. Methods for reducing biosensor membrane biofouling. Coll Surfaces B Biointerfaces. 2000;18(3–4):197–219.

    Article  CAS  Google Scholar 

  27. Chefer VI, Thompson AC, Zapata A, Shippenberg TS. Overview of brain microdialysis. Curr Protocols Neurosci. 2009;47(1):7–728.

    Article  Google Scholar 

  28. Müller M. Introduction to the microdialysis technology. In: Müller M, editor. Microdialysis in drug development. New York, NY: Springer New York; 2013. p. 3–12.

    Chapter  Google Scholar 

  29. Kendrick K. Microdialysis measurement of in vivo neoropeptide release. J Neurosci Methods. 1990;34(1–3):35–46.

    Article  CAS  PubMed  Google Scholar 

  30. Clough GF. Microdialysis of large molecules. The AAPS J. 2005;7(3):E686–92.

    Article  Google Scholar 

  31. Ao X, Sellati TJ, Stenken JA. Enhanced microdialysis relative recovery of inflammatory cytokines using antibody-coated microspheres analyzed by flow cytometry. Anal Chem. 2004;76(13):3777–84.

    Article  CAS  PubMed  Google Scholar 

  32. Kjellstrom S, Appels N, Ohlrogge M, Laurell T, Marko-Varga G. Microdialysis—a membrane based sampling technique for quantitative determination of proteins. Chromatographia. 1999;50(9-10):539–46.

    Article  CAS  Google Scholar 

  33. Pieber TR, Birngruber T, Bodenlenz M, Hoffere C, Selma M, Tiffner K, et al. Open flow microperfusion: an alternative method to microdialysis? In: Mueller M, editor. New York: Springer; 2012.

  34. Schaupp L, Ellmerer M, Brunner GA, Wutte A, Sendlhofer G, Trajanoski Z, et al. Direct access to interstitial fluid in adipose tissue in humans by use of open-flow microperfusion. Am J Physiol Endocrinol Metab. 1999;276(2):E401–8.

    Article  Google Scholar 

  35. van den Brink FTG, Phisonkunkasem T, Asthana A, Bomer JG, van den Maagdenberg AMJM, Tolner EA, et al. A miniaturized push–pull-perfusion probe for few-second sampling of neurotransmitters in the mouse brain. Lab Chip. 2019;19(8):1332–43.

    Article  PubMed  Google Scholar 

  36. Lee W, Ngernsutivorakul T, Mabrouk O, Wong J-M, Dugan C, Pappas S, et al. Microfabrication and in vivo performance of a microdialysis Probe with embedded membrane. Anal Chem. 2016;88.

  37. Queval A, Perrault CM, Qasaimeh MA, McKinney R, Juncker D, editors. Design and fabrication of a PDMS microfluidic probe and perfusion chamber for microfluidic experiments with organotypic brain slices. Micro total analysis systems. 2008.

  38. Trajanoski Z, Brunner GA, Schaupp L, Ellmerer M, Wach P, Pieber TR, et al. Open-flow microperfusion of subcutaneous adipose tissue for on-line continuous ex vivo measurement of glucose concentration. Diabetes Care. 1997;20(7):1114–21.

    Article  CAS  PubMed  Google Scholar 

  39. Cepeda DE, Hains L, Li D, Bull J, Lentz SI, Kennedy RT. Experimental evaluation and computational modeling of tissue damage from low-flow push–pull perfusion sampling in vivo. J Neurosci Methods. 2015;242:97–105.

    Article  PubMed  Google Scholar 

  40. Ellmerer M, Schaupp L, Brunner GA, Sendlhofer G, Wutte A, Wach P, et al. Measurement of interstitial albumin in human skeletal muscle and adipose tissue by open-flow microperfusion. Am J Physiol Endocrinol Metab. 2000;278(2):E352–6.

    Article  Google Scholar 

  41. Torto N, Gorton L, Laurell T, Marko-Varga G. Technical issues of in vitro microdialysissampling in bioprocess monitoring. Trends Anal Chem. 1999;18(4).

  42. Hsiao JK, Ball BA, Morrison PF, Mefford IN, Bungay PM. Effects of different semipermeable membranes on in vitro and in vivo performance of microdialysis probes. J Neurochem. 1990;54(4):1449–52.

    Article  CAS  PubMed  Google Scholar 

  43. Hammarlund-Udenaes M. Microdialysis as an important technique in systems pharmacology—a historical and methodological review. Aaps J. 2017;19(5):1294–303.

    Article  CAS  PubMed  Google Scholar 

  44. Levine J, Powell K. Microdialysis for measurement of neuroendocrine peptides. Methods Enzymol. 1989;169:166–81.

    Article  Google Scholar 

  45. Burns L, Van Alstine J, Harris J. Poly(ethylene glycol) grafted to quartz: analysis in terms of a site-dissociation model of electroosmotic fluid flow. Langmuir. 1995;11:2768–76.

    Article  CAS  Google Scholar 

  46. Clément R, Malinovsky J-M, Dollo G, Le Corre P, Chevanne F, Le Verge R. In vitro and in vivo microdialysis calibration using retrodialysis for the study of the cerebrospinal distribution of bupivacaine. J Pharmaceut Biomed Anal. 1998;17(4–5):665–70.

    Article  Google Scholar 

  47. Williams LR, Vahlsing HL, Lindamood T, Varon S, Gage FH, Manthorpe M. A small-gauge cannula device for continuous infusion of exogenous agents into the brain. Exp Neurol. 1987;95(3):743–54.

    Article  CAS  PubMed  Google Scholar 

  48. Cepeda DE. Brain tissue response in neurochemical sampling: microdialysis and low-flow push-pull perfusion. 2013;2013:78–101.

  49. Kottegoda S, Shaik I, Shippy SA. Demonstration of low flow push–pull perfusion. J Neurosci Methods. 2002;121(1):93–101.

    Article  PubMed  Google Scholar 

  50. De Lange EC, De Vries JD, Zurcher C, Danhof M, De Boer AG, Breimer DD. The use of intracerebral microdialysis for the determination of pharmacokinetic profiles of anticancer drugs in tumor-bearing rat brain. 1995.

  51. Bungay PM, Morrison PF, Dedrick RL. Steady-state theory for quantitative microdialysis of solutes and water in vivo and in vitro. Life Sci. 1990;46(2):105–19.

    Article  CAS  PubMed  Google Scholar 

  52. Menacherry S, Hubert W, Justice JB. In vivo calibration of microdialysis probes for exogenous compounds. Anal Chem. 1992;64(6):577–83.

    Article  CAS  PubMed  Google Scholar 

  53. Olson RJ, Justice JB. Quantitative microdialysis under transient conditions. Anal Chem. 1993;65(8):1017–22.

    Article  CAS  PubMed  Google Scholar 

  54. Jacobson I, Sandberg M, Hamberger A. Mass transfer in brain dialysis devices—a new method for the estimation of extracellular amino acids concentration. J Neurosci Methods. 1985;15(3):263–8.

    Article  CAS  PubMed  Google Scholar 

  55. Erickson M, Banks W. Cytokine and chemokine responses in serum and brain after singles and repeated injections of lipopolysaccharide: multiplex quantification with path analysis. Brain Behavior Immunity. 2011;25:449–55.

    Article  Google Scholar 

  56. Tsuji A, Tamai I. Blood–brain barrier function of P-glycoprotein. Adv Drug Deliv Rev. 1997;25(2):287–98.

    Article  CAS  Google Scholar 

  57. Szarowski DH, Andersen MD, Retterer S, Spence AJ, Isaacson M, Craighead HG, et al. Brain responses to micro-machined silicon devices. Brain Res. 2003;983(1–2):23–35.

    Article  CAS  PubMed  Google Scholar 

  58. Seymour JP, Kipke DR. Neural probe design for reduced tissue encapsulation in CNS. Biomaterials. 2007;28(25):3594–607.

    Article  CAS  PubMed  Google Scholar 

  59. Mercanzini A, Cheung K, Buhl DL, Boers M, Maillard A, Colin P, et al. Demonstration of cortical recording using novel flexible polymer neural probes. Sens Actuators A Phys. 2008;143(1):90–6.

    Article  CAS  Google Scholar 

  60. Biran R, Martin DC, Tresco PA. Neuronal cell loss accompanies the brain tissue response to chronically implanted silicon microelectrode arrays. Exp Neurol. 2005;195(1):115–26.

    Article  CAS  PubMed  Google Scholar 

  61. Fassbender K, Schneider S, Bertsch T, Schlueter D, Fatar M, Ragoschke A, et al. Temporal profile of release of interleukin-1β in neurotrauma. Neurosci Lett. 2000;284(3):135–8.

    Article  CAS  PubMed  Google Scholar 

  62. Woodroofe MN, Sarna GS, Wadhwa M, Hayes GM, Loughlin AJ, Tinker A, et al. Detection of interleukin-1 and interleukin-6 in adult rat brain, following mechanical injury, by in vivo microdialysis: evidence of a role for microglia in cytokine production. J Neuroimmunol. 1991;33(3):227–36.

    Article  CAS  PubMed  Google Scholar 

  63. Nagashima G, Suzuki R, Asai J-i, Fujimoto T. Immunohistochemical analysis of reactive astrocytes around glioblastoma: an immunohistochemical study of postmortem glioblastoma cases. Clin Neurol Neurosurg. 2002;104(2):125–31.

    Article  PubMed  Google Scholar 

  64. Hyvarinen T, Hagman S, Ristola M, Sukki L, Veijula K, Kreutzer J, et al. Co-stimulation with IL-1beta and TNF-alpha induces an inflammatory reactive astrocyte phenotype with neurosupportive characteristics in a human pluripotent stem cell model system. Sci Rep. 2019;9(1):16944.

    Article  PubMed  PubMed Central  Google Scholar 

  65. Reato D, Cammarota M, Parra LC, Carmignoto G. Computational model of neuron-astrocyte interactions during focal seizure generation. Front Comput Neurosci. 2012;6:81.

    Article  PubMed  PubMed Central  Google Scholar 

  66. Ventura R, Harris KM. Three-dimensional relationships between hippocampal synapses and astrocytes. J Neurosci. 1999;19(16):6897–906.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  67. Vezzani A, Moneta D, Richichi C, Aliprandi M, Burrows SJ, Ravizza T, et al. Functional role of inflammatory cytokines and antiinflammatory molecules in seizures and epileptogenesis. Epilepsia. 2002;43(s5):30–5.

    Article  CAS  PubMed  Google Scholar 

  68. Meunier N, Marion-Poll F, Lansky P, Rospars JP. Estimation of the individual firing frequencies of two neurons recorded with a single electrode. Chem Senses. 2003;28(8):671–9.

    Article  PubMed  Google Scholar 

  69. Haidarliu S, Shulz D, Ahissar E. A multi-electrode array for combined microiontophoresis and multiple single-unit recordings. J Neurosci Methods. 1995;56(2):125–31.

    Article  CAS  PubMed  Google Scholar 

  70. Gorgulho A, De Salles AAF, Frighetto L, Behnke E. Incidence of hemorrhage associated with electrophysiological studies performed using macroelectrodes and microelectrodes in functional neurosurgery. J Neurosurg. 2005;102(5):888–96.

    Article  PubMed  Google Scholar 

  71. Greenbaum A, Anava S, Ayali A, Shein M, David-Pur M, Ben-Jacob E, et al. One-to-one neuron–electrode interfacing. J Neurosci Methods. 2009;182(2):219–24.

    Article  PubMed  Google Scholar 

  72. Zhang CD,  Tian N, Jiang Q. Liu W, Huinan. Electrochemical deposition of conductive polymers onto magnesium microwires for neural electrode applications. J Biomed Mater Res. 2018;106(7):1887–95.

  73. Sohal HS, Jackson A, Jackson R, Clowry GJ, Vassilevski K, O’Neill A, et al. The sinusoidal probe: a new approach to improve electrode longevity. Front Neuroeng. 2014;7(10).

  74. Schwaerzle M, Pothof F, Paul O, Ruther P, editors. High-resolution neural depth probe with integrated 460 NM light emitting diode for optogenetic applications. In: 2015 Transducers—2015 18th International Conference on Solid-State Sensors, Actuators and Microsystems (transducers); 2015 21–25 June 2015.

  75. Liu Y, Vickers JA, Henry CS. Simple and sensitive electrode design for microchip electrophoresis/electrochemistry. Anal Chem. 2004;76(5):1513–7.

    Article  CAS  PubMed  Google Scholar 

  76. Arcot Desai S, Gutekunst C-A, Potter SM, Gross RE. Deep brain stimulation macroelectrodes compared to multiple microelectrodes in rat hippocampus. Front Neuroeng. 2014;7(16).

  77. Worrell GA, Gardner AB, Stead SM, Hu S, Goerss S, Cascino GJ, et al. High-frequency oscillations in human temporal lobe: simultaneous microwire and clinical macroelectrode recordings. Brain. 2008;131(4):928–37.

    Article  PubMed  Google Scholar 

  78. Gabay T, Ben-David M, Kalifa I, Sorkin R, Abrams ZeR, Ben-Jacob E, et al. Electro-chemical and biological properties of carbon nanotube based multi-electrode arrays. Nanotechnology. 2007;18(3):035201.

    Article  PubMed  Google Scholar 

  79. Alcaide M, Taylor A, Fjorback M, Zachar V, Pennisi CP. Boron-doped nanocrystalline diamond electrodes for neural interfaces: in vivo Biocompatibility Evaluation. Front Neurosci. 2016;10(87).

  80. Hong G, Lieber CM. Novel electrode technologies for neural recordings. Nat Rev Neurosci. 2019;20(6):330–45.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  81. Marg E, Adams JE. Indwelling multiple micro-electrodes in the brain. Electroencephalogr Clin Neurophysiol. 1967;23(3):277–80.

    Article  CAS  PubMed  Google Scholar 

  82. Lee K-K, He J, Singh A, Massia S, Ehteshami G, Kim B, et al. Polyimide-based intracortical neural implant with improved structural stiffness. J Micromech Microeng. 2003;14(1):32–7.

    Article  Google Scholar 

  83. Ceyssens F, Puers R. Insulation lifetime improvement of polyimide thin film neural implants. J Neural Eng. 2015;12(5):054001.

    Article  PubMed  Google Scholar 

  84. Karumbaiah L, Saxena T, Carlson D, Patil K, Patkar R, Gaupp EA, et al. Relationship between intracortical electrode design and chronic recording function. Biomaterials. 2013;34(33):8061–74.

    Article  CAS  PubMed  Google Scholar 

  85. Hariz MI, Fodstad H. Do microelectrode techniques increase accuracy or decrease risks in pallidotomy and deep brain stimulation? Stereotact Funct Neurosurg. 1999;72:157–69.

    Article  CAS  PubMed  Google Scholar 

  86. Winestone JS, Zaidel A, Bergman H, Israel Z. The use of macroelectrodes in recording cellular spiking activity. J Neurosci Methods. 2012;206(1):34–9.

    Article  PubMed  Google Scholar 

  87. Gimsa J, Habel B, Schreiber U, Rienen Uv, Strauss U, Gimsa U. Choosing electrodes for deep brain stimulation experiments–electrochemical considerations. J Neurosci Methods. 2005;142(2):251–65.

    Article  PubMed  Google Scholar 

  88. Kiran R, Rousseau L, Lissorgues G, Scorsone E, Bongrain A, Yvert B, et al. Multichannel boron doped nanocrystalline diamond ultramicroelectrode arrays: design, fabrication and characterization. Sensors. 2012;12(6):7669–81.

    Article  CAS  PubMed  Google Scholar 

  89. Yano T, Tryk, D.A., Hashimoto, K., Fujishima, A. Electrochemical behavior of highly conductive boron-doped diamond electrodes for oxygen reduction in alkaline solution. J Electrochem Soc. 1998;145(6).

  90. Swain GM, Ramesham R. The electrochemical activity of boron-doped polycrystalline diamond thin film electrodes. Anal Chem. 1993;65(4):345–51.

    Article  CAS  Google Scholar 

  91. Yang K-H, Narayan RJ. Biocompatibility and functionalization of diamond for neural applications. Curr Opin Biomed Eng. 2019;10:60–8.

    Article  Google Scholar 

  92. Arcot Desai S, Rolston J, Guo L, Potter S. Improving impedance of implantable microwire multi-electrode arrays by ultrasonic electroplating of durable platinum black. Front Neuroeng. 2010;3(5).

  93. Rao L, Zhou H, Li T, Li C, Duan YY. Polyethylene glycol-containing polyurethane hydrogel coatings for improving the biocompatibility of neural electrodes. Acta Biomater. 2012;8(6):2233–42.

    Article  CAS  PubMed  Google Scholar 

  94. Azemi E, Stauffer WR, Gostock MS, Lagenaur CF, Cui XT. Surface immobilization of neural adhesion molecule L1 for improving the biocompatibility of chronic neural probes: in vitro characterization. Acta Biomater. 2008;4(5):1208–17.

    Article  CAS  PubMed  Google Scholar 

  95. Ahemaiti A, Ainla A, Jeffries GDM, Wigström h, Jesorka A, Jardemark K. A multifunctional pipette for localized drug administration to brain slices. Biophys J. 2014;106(2):191a.

    Article  Google Scholar 

  96. Kehr J, Dechent P, Kato T, Ögren SO. Simultaneous determination of acetylcholine, choline and physostigmine in microdialysis samples from rat hippocampus by microbore liquid chromatography/electrochemistry on peroxidase redox polymer coated electrodes. J Neurosci Methods. 1998;83(2):143–50.

    Article  CAS  PubMed  Google Scholar 

  97. Okada M, Kawata Y, Mizuno K, Wada K, Kondo T, Kaneko S. Interaction between Ca2+, K+, carbamazepine and zonisamide on hippocampal extracellular glutamate monitored with a microdialysis electrode. Br J Pharmacol. 1998;124(6):1277–85.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  98. Albery J, Galley P. Inventors dialysis electrode device. 1995.

  99. Obrenovitch TP, Richards DA, Sarna GS, Symon L. Combined intracerebral microdialysis and electrophysiological recording: methodology and applications. Journal of Neuroscience Methods. 1993;47(1-2):139–45.

    Article  CAS  PubMed  Google Scholar 

  100. Murphy LJ. Reduction of interference response at a hydrogen peroxide detecting electrode using electropolymerized films of substituted naphthalenes. Anal Chem. 1998;70(14):2928–35.

    Article  CAS  Google Scholar 

  101. Murphy LJ, Galley PT. Measurement in vitro of human plasma glycerol with a hydrogen peroxide detecting microdialysis enzyme electrode. Anal Chem. 1994;66(23):4345–53.

    Article  CAS  PubMed  Google Scholar 

  102. Yamamoto BK, Pehek EA. A neurochemical heterogeneity of the rat striatum as measured by in vivo electrochemistry and microdialysis. Brain Res. 1990;506(2):236–42.

    Article  CAS  PubMed  Google Scholar 

  103. Berners MOM, Boutelle MG, Fillenz M. Online measurement of brain glutamate with an enzyme/polymer-coated tubular electrode. Anal Chem. 1994;66(13):2017–21.

    Article  CAS  PubMed  Google Scholar 

  104. Shi G, Lu J, Xu F, Sun W, Jin L, Yamamoto K, et al. Determination of glutathione in vivo by microdialysis using liquid chromatography with a cobalt hexacyanoferrate chemically modified electrode. Anal Chim Acta. 1999;391(3):307–13.

    Article  CAS  Google Scholar 

  105. Petit-Pierre G, Bertsch A, Renaud P. Neural probe combining microelectrodes and a droplet-based microdialysis collection system for high temporal resolution sampling. Lab Chip. 2016;16:917–24.

    Article  CAS  PubMed  Google Scholar 

  106. Kang J, Kadam SD, Elmore JS, Sullivan BJ, Valentine H, Malla AP, et al. Transcranial photoacoustic imaging of NMDA-evoked focal circuit dynamics in the rat hippocampus. J Neural Eng. 2020;17(2):025001.

    Article  PubMed  PubMed Central  Google Scholar 

  107. Vondrakova K, Novotny P, Kubova H, Posusta A, Boron J, Faberova V, et al. Electrographic seizures induced by activation of ETA and ETB receptors following intrahippocampal infusion of endothelin-1 in immature rats occur by different mechanisms. Exp Neurol. 2020;328:113255.

    Article  CAS  PubMed  Google Scholar 

  108. Shokry IM, Sinha V, Da Silva G, Park SB, Callanan JJ, Tao R. Comparison of electroencephalogram (EEG) response to MDPV versus the hallucinogenic drugs MK-801 and ketamine in rats. Exp Neurol. 2019;313:26–36.

    Article  CAS  PubMed  Google Scholar 

  109. Lelkes Z, Abdurakhmanova S, Porkka-Heiskanen T. Cholinergic basal forebrain structures are not essential for mediation of the arousing action of glutamate. J Sleep Res. 2018;27(4):e12605.

    Article  PubMed  Google Scholar 

  110. Kanamori K. Faster flux of neurotransmitter glutamate during seizure—evidence from 13 C-enrichment of extracellular glutamate in kainate rat model. PLoS One. 2017;12(4):e0174845.

    Article  PubMed  PubMed Central  Google Scholar 

  111. Morales-Villagran A, Pardo-Pena K, Medina-Ceja L, Lopez-Perez S. A microdialysis and enzymatic reactor sensing procedure for the simultaneous registration of online glutamate measurements at high temporal resolution during epileptiform activity. J Neurochem. 2016;139(5):886–96.

    Article  CAS  PubMed  Google Scholar 

  112. Soukupova M, Binaschi A, Falcicchia C, Palma E, Roncon P, Zucchini S, et al. Increased extracellular levels of glutamate in the hippocampus of chronically epileptic rats. Neuroscience. 2015;301:246–53.

    Article  CAS  PubMed  Google Scholar 

  113. Konno D, Nakayama R, Tsunoda M, Funatsu T, Ikegaya Y, Sasaki T. Collection of biochemical samples with brain-wide electrophysiological recordings from a freely moving rodent. J Pharmacol Sci. 2019;139(4):346–51.

    Article  CAS  PubMed  Google Scholar 

  114. Bourdon AK, Spano GM, Marshall W, Bellesi M, Tononi G, Serra PA, et al. Metabolomic analysis of mouse prefrontal cortex reveals upregulated analytes during wakefulness compared to sleep. Sci Rep. 2018;8(1):11225.

    Article  PubMed  PubMed Central  Google Scholar 

  115. Hiller A, Loeffler S, Haupt C, Litza M, Hofmann U, Moser A. Electrical high frequency stimulation of the caudate nucleus induces local GABA outflow in freely moving rats. J Neurosci Methods. 2007;159(2):286–90.

    Article  CAS  PubMed  Google Scholar 

  116. Tamano H, Nishio R, Shakushi Y, Sasaki M, Koike Y, Osawa M, et al. In vitro and in vivo physiology of low nanomolar concentrations of Zn(2+) in artificial cerebrospinal fluid. Sci Rep. 2017;7:42897.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  117. Tamano H, Nishio R, Takeda A. Involvement of intracellular Zn(2+) signaling in LTP at perforant pathway-CA1 pyramidal cell synapse. Hippocampus. 2017;27(7):777–83.

    Article  CAS  PubMed  Google Scholar 

  118. Potschka H, Baltes S, Fedrowitz M, Loscher W. Impact of seizure activity on free extracellular phenytoin concentrations in amygdala-kindled rats. Neuropharmacology. 2011;61(5-6):909–17.

    Article  CAS  PubMed  Google Scholar 

  119. Varatharajan R, Joseph K, Neto SC, Hofmann UG, Moser A, Tronnier V. Electrical high frequency stimulation modulates GABAergic activity in the nucleus accumbens of freely moving rats. Neurochem Int. 2015;90:255–60.

    Article  CAS  PubMed  Google Scholar 

  120. Schumacher A, Haegele M, Spyth J, Moser A. Electrical high frequency stimulation of the nucleus accumbens shell does not modulate depressive-like behavior in rats. Behav Brain Res. 2020;378:112277.

    Article  CAS  PubMed  Google Scholar 

  121. Soukupova M, Falcicchia C, Lovisari F, Ingusci S, Barbieri M, Zucchini S, et al. Microdialysis of excitatory amino acids during eeg recordings in freely moving rats. J Vis Exp. 2018(141).

  122. During MJ, Spencer DD. Extracellular hippocampal glutamate and spontaneous seizure in the conscious human brain. Lancet. 1993;341(8861):1607–10.

    Article  CAS  PubMed  Google Scholar 

  123. Fried I, Wilson CL, Maidment NT, Engel J, Behnke E, Fields TA, et al. Cerebral microdialysis combined with single-neuron and electroencephalographic recording in neurosurgical patients. J Neurosurg. 1999;91(4):697–705.

    Article  CAS  PubMed  Google Scholar 

  124. Vezzani A, Ungerstedt U, French ED, Schwarcz R. In vivo brain dialysis of amino acids and simultaneous EEG measurements following intrahippocampal quinolinic acid injection: evidence for a dissociation between neurochemical changes and seizures. J Neurochem. 1985;45(2):335–44.

    Article  CAS  PubMed  Google Scholar 

  125. John R, John MJ. Evaluation of the microdialysis electrode concept as a generic biosensing technology. Electroanalysis. 1998;10(16):1125–9.

    Article  CAS  Google Scholar 

  126. Walker MC, Galley PT, Errington ML, Shorvon SD, Jefferys JG. Ascorbate and glutamate release in the rat hippocampus after perforant path stimulation: a “dialysis electrode” study. J Neurochem. 1995;65(2):725–31.

    Article  CAS  PubMed  Google Scholar 

  127. Cavus I, Romanyshyn JC, Kennard JT, Farooque P, Williamson A, Eid T, et al. Elevated basal glutamate and unchanged glutamine and GABA in refractory epilepsy: microdialysis study of 79 patients at the yale epilepsy surgery program. Ann Neurol. 2016;80(1):35–45.

    Article  CAS  PubMed  Google Scholar 

  128. Mishra P, Lehmann J, Nair N. Inventors microdialysis probes and methods of use patent 5,441,481. 1994.

  129. Kucharczyk J, Gillies GT.  Inventors multi-probe system patent 6,626,902. 2003.

  130. Engström ER, Hillered L, Flink R, Kihlström L, Lindquist C, Nie JX, et al. Extracellular amino acid levels measured with intracerebral microdialysis in the model of posttraumatic epilepsy induced by intracortical iron injection. Epilepsy Res. 2001;43(2):135–44.

    Article  Google Scholar 

  131. Amberg G, Lindefors N. Intracerebral microdialysis: II. Mathematical studies of diffusion kinetics. J Pharmacolog Methods. 1989;22(3):157–83.

    Article  CAS  Google Scholar 

  132. Lopes da Silva F, Pijn JP, Boeijinga P. Interdependence of EEG signals: linear vs. nonlinear associations and the significance of time delays and phase shifts. Brain Topography. 1989;2(1):9–18.

    Article  CAS  PubMed  Google Scholar 

  133. Höller Y, Nardone R. Chapter Six - Quantitative EEG biomarkers for epilepsy and their relation to chemical biomarkers. In: Makowski GS, editor. Advances in clinical chemistry. 102: Elsevier; 2021; 271–336.

  134. Tang J, Zhang J, Ma J, Zhang G, Yang X. Astrocyte calcium wave induces seizure-like behavior in neuron network. Sci China Technolog Sci. 2017;60(7):1011–8.

    Article  CAS  Google Scholar 

  135. Milan J-L, Planell JA, Lacroix D. Computational modelling of the mechanical environment of osteogenesis within a polylactic acid–calcium phosphate glass scaffold. Biomaterials. 2009;30(25):4219–26.

    Article  CAS  PubMed  Google Scholar 

  136. Worrell GA, Jerbi K, Kobayashi K, Lina JM, Zelmann R, Le Van Quyen M. Recording and analysis techniques for high-frequency oscillations. Prog Neurobiol. 2012;98(3):265–78.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  137. Kendrick K. Use of microdialysis in neuroendocrinology. Methods Enzymol. 1989;168:182–205.

    Article  CAS  PubMed  Google Scholar 

  138. Shippenberg T, Thompson, A. Overview of microdialysis. Chapter 7, Unit 7.1 ed2001.

Download references

Acknowledgements

Not applicable.

Funding

This work was supported, in part, by Grant NS111756 “Peri-electrode large molecule microdialysis of the brain during induced seizures in mice, pigs, and humans with epilepsy undergoing resective surgery” from the National Institute of Neurological Disorders and Stroke awarded to CLH and GAW, and by Minnesota Partnership grant “Defining parameters and performance specifications for an implantable CNS drug testing device” awarded to TCB.

Author information

Authors and Affiliations

Authors

Contributions

All authors participated in the conception and writing of the manuscript. All authors provided critical editing and final approval. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Charles L. Howe.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Stangler, L.A., Kouzani, A., Bennet, K.E. et al. Microdialysis and microperfusion electrodes in neurologic disease monitoring. Fluids Barriers CNS 18, 52 (2021). https://doi.org/10.1186/s12987-021-00292-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12987-021-00292-x