Skip to main content
Fig. 2 | Fluids and Barriers of the CNS

Fig. 2

From: Measuring intracranial pressure by invasive, less invasive or non-invasive means: limitations and avenues for improvement

Fig. 2

Overview of wire-based and wireless methods for ICP monitoring. The image on the right shows that ICP is measured via a ventricular (V) catheter placed within the cerebral ventricles, and dedicated ICP sensors implanted within the brain parenchyma (P), or via the ICP sensor placed within the epidural (E) location. The invasive ICP source signals are transferred to a monitor that may reveal the ICP scores. For example, the ICP scores may be shown as numerical values, trend plots, or as the single ICP waves. The image on the left illustrates implantable sensors to the ventricles or parenchyma wherein the communication between sensor and external receiver is wireless. Illustration: Øystein Horgmo, University of Oslo

Back to article page