Skip to main content

Table 1 Hydrodynamic resistance of the opened shunt depends on the preset performance level.

From: Investigation of the hydrodynamic properties of a new MRI-resistant programmable hydrocephalus shunt

Performance level

Shunt without Catheter

Shunt with catheter (110 cm long, internal diameter 1.1 mm)

30 mmH2O

2.3 mmHg

1.58 (0.3)

4.61 (.71)

70 mmH2O

5.2 mmHg

2.06 (0.41)

5.12 (0.76)

110 mmH2O

8.1 mmHg

2.48 (0.52)

5.62 (0.82)

150 mmH2O

11.1 mmHg

5.18 (0.81)

8.21 (1.01)

200 mmH2O

14.7 mmHg

11.2 (1.31)

14.2 (1.42)

  1. Units are mmHg/ml/min; mean values and standard deviation (in parentheses). Each test was repeated 3 times in 3 valves.